Sub topic 13
Meter Bridge: Wire, Gaps, Jockey, Galvanometer and Resistance Box
What this chapter covers
This chapter describes the physical meter bridge: its parts, what each part does, and how to connect them. The balance-length formula comes in the next chapter, and this one only prepares the setup it uses.
You should already know these things:
- Ohm's law, and resistances in series and parallel.
- That resistance of a uniform wire is proportional to its length.
- That a galvanometer can act as a null detector (it shows whether current flows, not how much).
- The Wheatstone bridge balance condition from the previous chapter.
The meter bridge as a Wheatstone bridge
A meter bridge is a Wheatstone bridge built on a board. Two of the four arms are the two parts of one uniform wire, split at the point where a sliding contact touches it. The other two arms are a known resistance and the unknown resistance.
Because the wire is uniform, the resistance of any segment is proportional to its length. This lets you compare resistances by comparing lengths on a scale.
The parts and their jobs
| Part | Description | Job |
|---|---|---|
| Bridge wire | 1 m long, uniform, usually manganin or constantan, stretched along a metre scale | Forms two arms of the bridge. Its lengths act as a ratio of resistances. |
| Copper strips | Thick strips fixed to the board, with terminals | Join the wire ends and the gaps. Their resistance is negligible. |
| Left gap | Gap between the left L-shaped strip and the middle strip | Holds the known resistance R from the resistance box. |
| Right gap | Gap between the middle strip and the right L-shaped strip | Holds the unknown resistance X. |
| Jockey | Sliding contact with a knife edge | Touches the wire at a chosen point and divides it into two lengths. |
| Galvanometer | Connected between the middle strip and the jockey | Shows when no current flows, which is the balance point. |
| Resistance box | Box of standard resistances with plug keys | Provides a known, adjustable R. |
| Cell, key and optional rheostat | Connected across the two ends of the wire | Drive current through the circuit. |
The wire material matters. Manganin and constantan have a very small temperature coefficient of resistance. The wire heats slightly while current flows, so a material with a small coefficient keeps the wire's resistance nearly constant.
How the circuit is connected
Call the left end of the wire A, the right end B, the middle strip C and the jockey contact J.
- Connect the resistance box between A and C, in the left gap.
- Connect the unknown resistance X between C and B, in the right gap.
- Connect the cell, a one-way key and an optional rheostat across A and B, the two ends of the wire.
- Connect the galvanometer between C (the middle strip) and the jockey. Put a high series resistance with it as protection.
- Slide the jockey along the wire until the galvanometer shows zero deflection. The distance AJ is the balance length l.
The galvanometer is not in the main current path. It sits across the "bridge" between C and J, so at balance the potential at J equals the potential at C and no current passes through it.
Why each part is built this way
Thick copper strips
The strips must have almost zero resistance. If they did not, the resistance in each arm would include an unknown extra amount. The error is worst at the ends of the wire, where the strips join it. This is called an end error, and a later chapter covers how to reduce it.
A uniform wire
If the wire's cross-section varied, resistance would no longer be proportional to length, and the ratio l/(100 − l) would not equal the ratio of resistances. This is why the jockey must not scrape the wire. Scraping thins it locally and makes it non-uniform.
The jockey
The knife edge gives a single, narrow contact point so the length AJ is well defined. Press it lightly, lift it between trials, and never drag it along the wire. Dragging wears the wire and gives a false reading from a smeared contact.
The galvanometer as a null detector
You only need to find the point where the pointer does not move, so the galvanometer does not need calibration. A zero-centre galvanometer is convenient, because the direction of deflection tells you which side of the balance point you are on. The high series resistance protects it when the bridge is far from balance and large current would flow through it.
The resistance box
In a plug-type box, each plug short-circuits one coil. To insert a resistance, you remove the plug, and the resistance in the circuit is the sum of the values of the removed plugs. Loose plugs add contact resistance, so tighten every plug that should be in place. Always remove a plug before you start, so that R is not zero.
Choosing R for a good reading
Adjust R until the balance point falls between about 40 cm and 60 cm. Near the middle of the wire, a small shift of the jockey changes the galvanometer deflection by a larger amount, so the null point is sharper. The fractional error in the ratio l/(100 − l) is also smallest near l = 50 cm.
The code below computes the balance length you should expect for a trial R. It uses the balance condition R/X = l/(100 − l), which gives l = 100R/(R + X). Use it to check whether a trial value of R will put the balance point in the useful range.
def expected_balance_cm(known_r, unknown_x):
# Balance: R/X = l/(100 - l) -> l = 100 R / (R + X)
return 100.0 * known_r / (known_r + unknown_x)
x = 15.0 # approximate unknown resistance in ohm
for r in (2.0, 5.0, 10.0, 15.0, 20.0, 50.0):
l = expected_balance_cm(r, x)
ok = "good" if 40 <= l <= 60 else "poor"
print(f"R = {r:5.1f} ohm -> l = {l:5.1f} cm ({ok})")For X near 15 Ω, only R between about 10 Ω and 22 Ω keeps l inside 40–60 cm. The other values give a poor balance point.
Common setup mistakes
- Battery current left on. A continuous current heats the wire and changes its resistance. Press the key only while taking a reading.
- Unknown in the wrong gap. The formula in the next chapter assumes R in the left gap and X in the right gap. Swapping them inverts the ratio.
- Rusty or dirty terminals. Clean the connecting wire ends with sandpaper so contacts add no resistance.
- Galvanometer without protection. Without a series resistance, a large deflection can damage it.
- Reading l from the wrong end. The balance length is measured from the end next to the known resistance, which is the left end A.
Exercise
In a meter bridge a student places the resistance box in the left gap and the unknown resistance in the right gap. The galvanometer is connected between the middle strip and the jockey. (a) Why is the galvanometer not placed in series with the cell? (b) The student sees a deflection to the right at l = 30 cm and to the left at l = 70 cm. In which range should the balance point lie?
Show answer
(a) The galvanometer is a null detector. Its job is to compare the potential at the middle strip with the potential at the jockey. That is only possible when it joins those two points, so it must sit in the bridge arm and not in the cell circuit. (b) The deflection changes direction between 30 cm and 70 cm, so the balance point lies between those two readings. The student should move the jockey in smaller steps within that range until the pointer shows zero deflection.
Exercise
A student gets a balance point at l = 85 cm. What should they change, and why?
Show answer
A balance point near one end is a poor reading, because the ratio l/(100 − l) is then very sensitive to small errors and end errors are larger. The student should change R, bringing it closer to the unknown resistance, until the balance point is between 40 cm and 60 cm.