Sub topic 3
What Makes a Measurement System Good: Location and Spread
A measurement system is good when its readings are close to the truth, consistent with one another, and stable over time. This chapter assumes you know the mean, variance, and standard deviation, understand a normal distribution, and can read a run chart or control chart.
The two questions to ask
Every measurement system has two basic properties:
- Location describes where the readings are centered.
- Spread describes how much the readings vary.
Suppose a reference standard has a true length of 50.00 mm. If a gage repeatedly reports values near 50.00 mm, its location is good. If those readings differ only slightly, its spread is good.
A useful measurement model is:
\[ \text{Observed reading} = \text{true value} + \text{systematic error} + \text{random error} \]
Systematic error moves readings in a consistent direction. Random error causes readings to scatter.
| Property | Main question | Typical evidence |
|---|---|---|
| Accuracy | Is the average close to the reference value? | Bias |
| Precision | Do repeated readings stay close together? | Standard deviation |
| Stability | Does performance remain consistent over time? | Control chart |
| Resolution | Can the system detect practically important differences? | Smallest readable increment |
Accuracy and precision are different. A gage can be precise but inaccurate if it gives nearly identical readings that are all too high. It can be accurate on average but imprecise if individual readings scatter widely.
Location: bias and linearity
Bias is the difference between the average observed measurement and a known reference value:
\[ \text{Bias} = \bar{x} - x_{\text{reference}} \]
For example, if a reference part is 25.00 mm and ten readings average 25.08 mm, the estimated bias is +0.08 mm. The plus sign indicates that the system reads high.
Bias should be evaluated at the measurement value that matters. A constant offset may be acceptable in one application and unacceptable in another, depending on the tolerance.
Linearity describes whether bias changes across the operating range. A gage might read correctly near 10 mm but increasingly high near 100 mm. In that case, one correction cannot fix the whole range. Test several reference values across the expected measurement range and compare the bias at each value.
Spread: repeatability and reproducibility
Repeatability is the variation when the same operator measures the same item with the same equipment under the same conditions. It mainly reflects the equipment's short-term measurement error.
Reproducibility is the variation between operators, appraisers, or other measurement conditions. It shows whether different people obtain comparable results.
Together, repeatability and reproducibility are called gage R&R. A system may have low repeatability error but high reproducibility error if operators use different techniques. It may also have consistent operator behavior but poor repeatability because the instrument is unstable or difficult to use.
The goal is not zero variation. Every practical measurement system has some. The goal is for measurement variation to be small compared with the part or process variation that you need to detect.
A practical review sequence
Review a measurement system in this order:
- Confirm that the reference value and calibration method are appropriate.
- Check resolution. The instrument should show differences smaller than the change you need to detect.
- Check location with a reference standard. Estimate bias across the operating range.
- Check spread through repeated measurements.
- Check operator effects with more than one qualified operator.
- Check stability by plotting measurements over time.
A control chart can reveal drift, sudden shifts, or unusual cycles even when the overall average appears acceptable.
Exercise
Exercise
A reference part is 100.00 mm. An instrument gives ten readings with an average of 100.03 mm and a standard deviation of 0.01 mm. What do these numbers suggest?
Show answer
The estimated bias is +0.03 mm, so the instrument reads high on average. The small standard deviation suggests good repeatability for this reference, but the system still needs testing across its range, across operators, and over time.