Sub topic 12
Measuring resistance with a meter bridge
Prerequisites and scope
This chapter assumes you know the Wheatstone bridge balance condition P/Q = R/S, resistivity (R = ρL/A), and series and parallel combinations. The meter bridge is a Wheatstone bridge built so that the ratio of two arms is read off as a length ratio on a ruler. You will learn the working model, the formulas, the usual sources of error, and the numerical patterns JEE asks.
The apparatus
A meter bridge (also called a slide-wire bridge) has these parts:
- A uniform wire AC, exactly 100 cm long, stretched along a metre scale. It is usually made of manganin or constantan, which have a very small temperature coefficient of resistance.
- A thick copper strip along the top with two gaps. The left gap holds a known resistance R from a resistance box. The right gap holds the unknown resistance X.
- A cell with a key connected across A and C.
- A galvanometer connected between the middle point B of the copper strip (the junction between the two gaps) and a sliding jockey. The jockey touches the wire at a point D.
The four arms of the equivalent Wheatstone bridge are:
| Arm | Resistance |
|---|---|
| Known resistance, left gap | R |
| Unknown resistance, right gap | X |
| Wire from A to D (length l) | ρl/A |
| Wire from D to C (length 100 − l) | ρ(100 − l)/A |
The copper strip has negligible resistance, so R and X are effectively in series between A and C, and the two wire segments are in series between the same two points. The galvanometer bridges the two midpoints.
Deriving the balance condition
Slide the jockey along the wire until the galvanometer shows zero deflection. At that point no current flows through the galvanometer, so B and D are at the same potential. The Wheatstone balance condition gives
R / X = (resistance of AD) / (resistance of DC).
Because the wire is uniform, resistance is proportional to length. If AD has length l cm, then DC has length (100 − l) cm, and the ratio of the wire resistances equals the ratio of lengths:
R / X = l / (100 − l)
so
X = R (100 − l) / l.
Notice that the unknown resistance sits opposite the longer-to-shorter ratio. If the balance point is near the middle (l close to 50 cm), then X is close to R. This observation drives the best practice described below.
The following code implements the formula and checks a few cases.
def meter_bridge_unknown(R, l):
# R in the left gap, unknown X in the right gap, balance length l cm from the left end
return R * (100 - l) / l
# Balance at 40 cm with R = 10 ohm
print(meter_bridge_unknown(10.0, 40.0)) # 15.0 ohm
# Balance at the midpoint means X equals R
print(meter_bridge_unknown(10.0, 50.0)) # 10.0 ohm
# Balance at 25 cm means X is three times R
print(meter_bridge_unknown(10.0, 25.0)) # 30.0 ohmWorked example
A meter bridge has R = 10 Ω in the left gap. The jockey gives zero deflection at l = 40 cm.
- The wire segments have lengths 40 cm and 60 cm, so their resistance ratio is 40 : 60.
- R / X = 40 / 60, so X = 10 × 60 / 40 = 15 Ω.
Now suppose X is made of a wire of length 1.5 m and diameter 0.4 mm. Its resistivity follows from X = ρL/A. With A = π(0.2 × 10⁻³)² ≈ 1.257 × 10⁻⁷ m²,
ρ = X A / L = 15 × 1.257 × 10⁻⁷ / 1.5 ≈ 1.26 × 10⁻⁶ Ω·m.
This two-step pattern, a bridge reading followed by a resistivity calculation, is common in JEE practical-based questions.
Why the balance point should be near the middle
The sensitivity of the measurement depends on where the balance falls. Take the percentage error in X caused by a small error Δl in reading l. From X = R(100 − l)/l,
ΔX / X = Δl × [1/l + 1/(100 − l)] = Δl × 100 / [l(100 − l)].
The quantity l(100 − l) is largest at l = 50, so the relative error is smallest there. The practical rules are:
- Choose R so that the balance point lands between about 40 cm and 60 cm.
- Near the ends of the wire, a 1 mm reading error becomes a much larger percentage error in X.
- If the first trial balances near an end, change R by a factor close to the ratio found and try again.
The table shows how the relative error scales for a fixed Δl = 0.1 cm.
| l (cm) | l(100 − l) | ΔX/X |
|---|---|---|
| 10 | 900 | 1.1 % |
| 30 | 2100 | 0.48 % |
| 50 | 2500 | 0.40 % |
| 80 | 1600 | 0.63 % |
Procedure and good practice
- Connect the circuit with the key open. Put a suitable R in the left gap and the unknown X in the right gap.
- Close the key, and tap the jockey gently on the wire. Do not drag it, because dragging scrapes the wire and changes its cross-section.
- Find the point where the galvanometer reads zero. Record l.
- Interchange R and X, keeping the cell and wire unchanged, and find the new balance length l′ measured from the same end. This helps cancel end corrections (see below).
- Repeat with several values of R and average the computed X.
- Keep the key closed only briefly, so that the wire does not heat and change resistance.
A protective resistor is often placed in series with the galvanometer. It limits the current when the jockey is far from balance, and it is shorted out for the final fine adjustment.
Sources of error
- End resistances. The wire does not start exactly at the scale zero. The copper strips and the contacts at A and C add small extra resistances α and β to each end. The true condition is then R / X = (l + α) / (100 − l + β). Interchanging R and X gives a second equation, and combining the two readings removes the end correction from the result.
- Non-uniform wire. If the cross-section varies, resistance is no longer proportional to length, and the length ratio no longer equals the resistance ratio.
- Heating. Current through the wire raises its temperature. The wire material should have a low temperature coefficient, and the current should flow only while taking a reading.
- Contact resistance. Loose connections and oxidised surfaces add unknown resistances to the gaps. Clean and tighten them.
- Parallax and jockey pressure. Reading the scale at an angle, or pressing hard enough to bend the wire, changes the measured l.
For the interchange method, let the first balance be at l and the second at l′ after swapping. With end corrections, the two equations are
R / X = (l + α) / (100 − l + β) and X / R = (l′ + α) / (100 − l′ + β).
When α = β = 0, the product of the two ratios is 1, which gives l(l′) = (100 − l)(100 − l′), that is, l + l′ = 100. Any departure from l + l′ = 100 signals end corrections.
def meter_bridge_with_swap(R, l, l_swapped):
# Check whether the two readings are consistent with zero end correction
ideal = abs((l + l_swapped) - 100) < 0.2
X = R * (100 - l) / l
return X, ideal
print(meter_bridge_with_swap(10.0, 40.0, 60.0)) # (15.0, True)
print(meter_bridge_with_swap(10.0, 40.0, 58.0)) # (15.0, False)Common mistakes
- Using the wrong segment in the ratio. R is on the same side as the wire length l. Mixing up which gap holds R changes the answer from R(100 − l)/l to R l/(100 − l).
- Treating l as the distance from the right end. Always note which end the length is measured from.
- Forgetting that the formula needs R and X in series with each other and the two wire parts in series with each other. If a resistor is placed in parallel with X, you must first find its equivalent with X and then apply the condition.
- Assuming that a larger R always improves accuracy. It only helps if it moves the balance point toward 50 cm.
- Confusing balance at both ends with a valid reading. A zero deflection at l = 0 or l = 100 means that the circuit is open or shorted, not that the bridge is balanced.
Typical JEE variations
Combination in a gap. If the right gap holds two resistors X₁ and X₂ in parallel, the balance gives R / Xeq = l / (100 − l), with Xeq = X₁X₂ / (X₁ + X₂). Measuring X_eq lets you recover X₂ if X₁ is known.
Shifted balance after a change. If a resistor is added in parallel with R, the effective left resistance falls, and the balance point moves toward A. Work out the new R_eq and use the same formula. The direction of the shift is a quick check on your answer: a smaller left resistance requires a shorter left wire segment.
Wire of different material. Because only the ratio of lengths appears, the resistivity of the bridge wire cancels, so the formula holds for any uniform wire.
Unequal gap resistances. If the cell's internal resistance changes, the balance point does not move. At balance no current flows through the galvanometer, so the position depends only on the resistance ratio, not on the EMF or internal resistance of the supply.
Exercise
In a meter bridge, the known resistance in the left gap is 12 Ω and the unknown resistance X is in the right gap. The balance point is at 30 cm from the left end. (a) Find X. (b) When a 4 Ω resistor is connected in parallel with X, where does the new balance point lie, measured from the left end?
Show answer
(a) X = R(100 − l)/l = 12 × 70 / 30 = 28 Ω. (b) The parallel combination is 28 × 4 / (28 + 4) = 3.5 Ω. The balance length l satisfies 12 / 3.5 = l / (100 − l), so 12(100 − l) = 3.5 l, which gives 1200 = 15.5 l and l ≈ 77.4 cm. The balance point moves toward the right end because the right gap resistance has fallen.
Exercise
A student finds the balance point at 20 cm with R = 8 Ω. She wants a more accurate result. What value of R should she choose so that the balance point comes close to 50 cm, and why does this improve accuracy?
Show answer
From the first reading, X = 8 × 80 / 20 = 32 Ω. For balance at 50 cm, R should equal X, so R = 32 Ω. At l = 50 cm the product l(100 − l) is at its maximum of 2500, so a fixed reading error Δl gives the smallest relative error ΔX/X = 100Δl / [l(100 − l)].
Exercise
In a meter bridge, balance is found at l = 45 cm. After the two resistances are interchanged, balance is found at l′ = 54 cm. Do the readings suggest end corrections, and why?
Show answer
Without end corrections the two balance lengths would add to 100 cm. Here l + l′ = 99 cm, so there is a small departure of 1 cm. It suggests small end resistances at A and C, equal to a few millimetres of wire, which is why the interchange method and averaging are used.