Sub topic 12
What to do when the measurement system fails
This chapter assumes you can run a crossed gage R&R study and read its output: the variance components, %study variation, %tolerance and ndc, plus the standard graphs. The earlier chapters cover those topics. Here you will decide what to change when the results say the measurement system is not adequate.
Diagnose before you act
A failed study tells you the measurement system is too noisy for its purpose. It does not tell you why. Fixing the wrong source wastes money. Retraining operators does nothing if the gage has poor resolution.
Start with the variance components and ask which one dominates the gage R&R:
- Repeatability is the variation when one operator measures the same part repeatedly with the same gage. It points at the gage, the fixture, the method or the environment.
- Reproducibility is the variation between operators. It points at training, procedure, fixturing technique or operator-dependent judgment.
- Operator-by-part interaction means operators disagree on some parts but not others. It points at features that are hard to locate or read.
- Part-to-part variation that is small relative to the measurement error can make a good gage look bad.
Then look at the graphs before choosing a fix. A range chart with many ranges of zero or a few repeated values suggests inadequate resolution. A range chart with one operator's ranges consistently higher suggests a technique problem for that operator. Points on the by-operator plot that sit systematically apart suggest a reproducibility problem, and a crossing line on the operator-by-part plot suggests interaction.
Check the study before blaming the gage
Two study mistakes produce a failing result from a gage that is fine.
- Parts too similar. %contribution and %study variation compare measurement variance with total variance, which includes part-to-part variance. If the study parts come from a narrow band, part variance is understated, and the ratio looks worse than it would for the real process. Parts should span the range you will see in production, ideally including some near the specification limits.
- Parts too spread out. If you deliberately chose extreme parts, part variance is overstated, and a poor gage looks better than it is. This is why %tolerance matters alongside %study variation. It does not depend on the sample of parts.
If you suspect a sampling problem, rerun the study with better parts before spending on the gage. Also check for study conduct problems: operators who could see earlier readings, parts measured in a non-random order, or a gage that changed during the study. Randomize the run order and blind the operators.
Match the fix to the source
| Symptom | Likely cause | Typical actions |
|---|---|---|
| Large repeatability, ranges of zero or few distinct values | Resolution too coarse | Use a gage with finer discrimination. A common rule is that the gage increment should be no more than one tenth of the tolerance or process spread. |
| Large repeatability, ranges erratic | Worn or loose gage, unstable fixture, poor lighting, vibration, temperature | Maintain or repair the gage, redesign the fixture, control the environment |
| Large reproducibility, one operator differs | Technique or training | Write a clear work instruction, train and certify operators, add a visual aid |
| Large reproducibility, all operators differ | Ambiguous method or unclear definition of the feature | Define where and how to measure, and add a locating fixture |
| Significant interaction | Some parts are hard to measure, such as burrs, taper or soft material | Define the measurement point, change the fixture, or use a different method for those features |
| Bias or drift found in a separate study | Calibration or stability problem | Recalibrate, shorten the calibration interval, use a master part and plot it on a control chart |
Resolution deserves special attention because it caps what any other fix can achieve. If the gage reads in steps that are coarse relative to the process spread, rounding is a built-in error source. No amount of operator training will remove it.
Fix one source at a time where you can, and record what you changed. If you change the gage, the fixture and the procedure together and the study passes, you will not know which change mattered, and you will not know what to protect when the process is later modified.
Predict the effect before you spend
Variance components add, so you can estimate the result of a fix before you make it. Suppose an improvement halves the repeatability variance, or training halves the reproducibility variance. Or suppose the process will use the average of several readings, which divides the repeatability variance by the number of readings, n.
The following code projects %study variation and ndc for these scenarios. It uses variance components from a failed study.
def project(var_rep, var_repro, var_part,
n_avg=1, rep_scale=1.0, repro_scale=1.0):
"""Project gage R&R metrics after a change.
n_avg: readings averaged per reported value (reduces repeatability only).
rep_scale, repro_scale: multipliers on the variance components."""
rep = var_rep * rep_scale / n_avg
repro = var_repro * repro_scale
grr = rep + repro
total = grr + var_part
return {
"pct_study_var": round(100 * (grr / total) ** 0.5, 1),
"ndc": int(1.41 * (var_part / grr) ** 0.5),
}
var_rep, var_repro, var_part = 0.08, 0.03, 1.0
scenarios = {
"baseline": {},
"average 4 readings": {"n_avg": 4},
"halve reproducibility variance": {"repro_scale": 0.5},
"halve repeatability variance": {"rep_scale": 0.5},
"average 4 and halve reproducibility": {"n_avg": 4, "repro_scale": 0.5},
}
for name, kwargs in scenarios.items():
print(name, project(var_rep, var_repro, var_part, **kwargs))The baseline gives about 31.5% study variation and 4 distinct categories. That is marginal at best and fails a strict criterion. Averaging four readings brings it to about 21.8% and 6 categories. Halving the reproducibility variance alone helps less, about 29.5% and still 4 categories, because repeatability is the larger component here. This is the point of projecting first: the largest component is the one worth attacking.
Interim measures while the fix is built
A fix can take weeks. Meanwhile the process still needs decisions.
- Average repeated readings. This works only for the repeatability part, as the code showed. It also costs time, and the readings must be real re-measurements, including repositioning the part. Repeating a display reading proves nothing.
- Use guard bands. Tighten the acceptance limits inside the specification limits by an amount tied to the measurement uncertainty. This lowers the chance of shipping a bad part and raises the chance of rejecting a good one, so it moves cost from the customer to you.
- Remeasure borderline parts. Parts that read near a limit are the ones whose classification is most likely wrong. Route only those to a better instrument.
- Restrict who measures. If one operator or one gage is clearly better, use it temporarily. Document this as a containment, not a solution.
Decide how much improvement is enough
Acceptance guidelines are rules of thumb, not laws. A common reading treats under 10% study variation as good, 10% to 30% as conditionally acceptable, and over 30% as unacceptable. For ndc, 5 or more is usually the minimum. The right decision depends on what the measurement is used for.
- Controlling a process or improving it: you care about distinguishing the process's own variation, so %study variation and ndc matter most.
- Accepting or rejecting product near a limit: %tolerance matters most, and the cost of a wrong decision sets how much error you can tolerate.
- Safety-critical characteristics: demand the lower end of the range.
A marginal gage may be acceptable if replacing it costs more than the damage its errors cause, and if the customer agrees. Record the reasoning so the decision can be reviewed.
Verify and sustain the improvement
After the corrective action, rerun the study with the same design and the same kind of parts, so the numbers are comparable. Compare variance components, not just the final percentages. The component you targeted should drop, and the others should not rise. If they do, the fix introduced a new problem, such as a new fixture that is harder to use consistently.
Then protect the gain. Update the work instruction and the control plan, set a calibration and re-verification schedule, and track a master part over time to catch drift. A measurement system that passes once can degrade through wear, staff turnover or a change in the product.
Exercise
A gage R&R study gives variance components of 0.05 for repeatability, 0.002 for reproducibility and 0.6 for part-to-part variation. The study fails the %study variation criterion. Which corrective action should you consider first, and why is retraining operators a poor first choice?
Show answer
Repeatability is by far the dominant component, so look at the gage, fixture, resolution and environment first. Gage R&R variance is 0.052 and reproducibility contributes only 0.002 of it, about 4%. Retraining operators can at best remove that small share, so it cannot bring the study into an acceptable range. A finer gage, a better fixture or averaging several readings attacks the 0.05 term.
Exercise
Using the projection model in this chapter, a process averages 4 readings per reported value. Repeatability variance is 0.08, reproducibility variance is 0.03 and part variance is 1.0. What is the projected gage R&R variance, and which component remains unchanged by the averaging?
Show answer
Repeatability becomes 0.08 / 4 = 0.02, so gage R&R variance is 0.02 + 0.03 = 0.05. The reproducibility variance of 0.03 is unchanged, because averaging repeated readings by the same operator does not reduce differences between operators.