Sub topic 3
What Makes a Measurement System Good: Location and Spread
This chapter assumes you know the mean, variance and standard deviation, and that you can read a run chart. It also assumes you have seen the split of observed variation into part variation and measurement variation, which the previous chapter covers. Here we name the properties that decide whether a measurement system is good. The following chapters study each one in detail.
Two questions about every reading
A measurement system is everything that produces a number: the gage, the fixture, the operator, the procedure, the software and the environment. When it measures the same part repeatedly, it produces a distribution of readings. That distribution has a center and a width, and each one answers a different question.
- Location: Is the center of the readings in the right place? A shifted center means the system reads systematically high or low.
- Spread: How much do the readings scatter around their own center? Wide scatter means any single reading is unreliable, even when the center is right.
A third property, resolution, limits how finely the system can express a value at all.
Location and spread are independent. A system can be centered and noisy, or tight and shifted. Good systems need both properties in order.
Location: is the center right?
Location properties compare readings with a reference value. A reference value is the best available estimate of the true value of a part. It usually comes from a calibrated master, a certified standard or a measurement by a higher-accuracy lab. You never know the true value exactly, so a reference value always carries its own small uncertainty.
Bias
Bias is the difference between the average of repeated readings and the reference value.
bias = (mean of readings) − (reference value)
A positive bias means the system reads high. If a 10.00 mm master reads 10.05 mm on average, the bias is +0.05 mm. Bias at one point is the simplest location check. Calibration exists to find and remove it.
Linearity
Linearity is how bias changes across the operating range of the gage. A caliper may read correctly near 5 mm and read 0.1 mm high near 50 mm. A single-point bias study would miss this. You detect linearity problems by measuring several references that span the range and checking whether bias trends with size.
Stability
Stability is how bias changes over time. A gage that was correct on Monday may drift by Friday because of wear, temperature, or a loose fixture. You check stability by measuring the same master at regular intervals and plotting the results on a control chart. A stable system shows only random variation around a constant mean.
The next chapter covers how to run and analyze all three studies. For now, remember that bias, linearity and stability are all one idea. Each describes how far the center of the readings is from the truth, at one point, across sizes, or across time.
Spread: how much do readings scatter?
Spread is a property of the measurement system itself, so it is measured with the standard deviation of repeated readings. It has two sources that you separate on purpose.
- Repeatability is the variation when one operator measures the same part many times with the same gage under the same conditions. It is the within-condition scatter. It is sometimes called equipment variation, though the part, the method and the environment also contribute.
- Reproducibility is the variation caused by changing a condition: usually the operator, but also the gage, the day, the lab or the fixture setup. It shows up as differences between the average readings of each condition.
The two combine as variances, not as standard deviations:
σ²(measurement) = σ²(repeatability) + σ²(reproducibility)
This addition rule matters. If repeatability has a standard deviation of 0.03 and reproducibility has 0.04, the combined standard deviation is √(0.03² + 0.04²) = 0.05, not 0.07.
A later chapter covers repeatability and reproducibility fully, and the gage R&R chapters estimate them. The point here is classification. Spread errors are random from reading to reading, so averaging more readings reduces them. Location errors are systematic, so averaging never removes them. Only calibration and correction do.
Resolution
Resolution, also called discrimination, is the smallest change the system can display or detect. A digital scale that shows 0.1 g cannot tell a 10.02 g part from a 10.04 g part. Both read 10.0.
Coarse resolution hides real variation and it distorts spread estimates. Rounded readings pile up on a few values, so ranges between repeated readings become zero or one increment. The measurement system then looks more repeatable than it really is.
Two practical checks are common:
- Rule of ten: the smallest increment should be no larger than about one tenth of the tolerance or of the process variation you want to see. Some guides allow a looser limit, but a tenth is the usual target.
- Range chart check: on a range chart from a gage study, if the plotted ranges take only about three or fewer distinct values, resolution is too coarse for the job.
Resolution is necessary but not sufficient. A gage with many digits can still be biased and noisy. The digits show how finely the display counts, not how accurate the number is.
A short simulation
This script simulates a gage with a known bias and known spread, then estimates both from repeated readings of a reference part. It then rounds the same readings to a coarse increment to show the effect of resolution.
import numpy as np
from scipy import stats
rng = np.random.default_rng(7)
reference = 10.00 # reference value of the master
true_bias = 0.05 # the gage reads 0.05 high
sd_gage = 0.08 # repeatability standard deviation
n = 25
readings = reference + true_bias + rng.normal(0, sd_gage, n)
def location_and_spread(x, reference):
n = len(x)
s = x.std(ddof=1)
bias = x.mean() - reference
se = s / np.sqrt(n)
t = stats.t.ppf(0.975, n - 1)
return bias, (bias - t * se, bias + t * se), s
bias, ci, s = location_and_spread(readings, reference)
print(f"bias = {bias:.3f}, 95% CI = ({ci[0]:.3f}, {ci[1]:.3f}), sd = {s:.3f}")
print("bias significant:", not (ci[0] <= 0 <= ci[1]))
for step in (0.01, 0.1):
rounded = np.round(readings / step) * step
distinct = len(np.unique(np.round(rounded, 6)))
print(f"increment {step}: {distinct} distinct values, "
f"sd = {rounded.std(ddof=1):.3f}")Read the output in three parts. The confidence interval for the bias tells you whether the center differs from the reference by more than chance. The standard deviation estimates repeatability for this one operator and gage. The last loop shows resolution: at the 0.1 increment, the 25 readings collapse onto a handful of values, and the standard deviation no longer describes the true noise.
Common errors
- Confusing accuracy with precision. Accuracy, often called trueness for the average of readings, concerns location. Precision concerns spread. A tight cluster of readings far from the reference is precise but biased.
- Trusting display digits. More digits do not mean a smaller error. Check the real spread with repeated readings.
- Adding standard deviations. Add variances, then take the square root.
- Assuming averaging fixes everything. Averaging reduces random spread. It leaves bias and drift untouched.
- Using an unverified reference. If the master itself is off, every bias estimate inherits that error. Use certified references with a known uncertainty.
- Checking location once. One calibration shows the bias at one moment and one size. Linearity and stability need separate checks.
Summary table
| Property | Question it answers | Type of error | Typical remedy |
|---|---|---|---|
| Bias | Is the average at one point right? | Systematic | Calibration, correction |
| Linearity | Does bias change with size? | Systematic | Calibrate across the range, replace the gage |
| Stability | Does bias change over time? | Systematic | Maintenance, regular checks, control chart |
| Repeatability | How much do repeat readings scatter? | Random | Better gage, fixture or method |
| Reproducibility | How much do conditions differ? | Random between groups | Training, standard procedure |
| Resolution | How fine are the readings? | Quantization | Gage with a smaller increment |
Exercise
A gage has a repeatability standard deviation of 0.06 mm and an operator-to-operator (reproducibility) standard deviation of 0.08 mm. What is the combined measurement standard deviation, and is averaging five readings enough to remove a consistent +0.10 mm offset?
Show answer
Add variances: 0.06² + 0.08² = 0.0036 + 0.0064 = 0.0100, so the combined standard deviation is √0.0100 = 0.10 mm. Averaging five readings reduces random spread but does not remove the +0.10 mm offset. That offset is bias, a systematic location error, and only calibration or a correction removes it.
Exercise
A process has a tolerance of 0.5 mm. A micrometer displays readings in steps of 0.05 mm. Does it meet the rule of ten, and what would you check on a gage-study range chart to confirm your answer?
Show answer
One tenth of the tolerance is 0.05 mm, so the 0.05 mm increment just meets the rule of ten. It is at the limit and leaves no margin. On the range chart, count how many distinct range values appear. If only about three or fewer occur, the resolution is too coarse for the job, and a finer increment is needed.