Sub topic 10
Reading the results against acceptance guidelines
This chapter assumes you can already compute variance components for a crossed gage R&R study with the ANOVA method and derive %contribution, %study variation, %tolerance and ndc from them. The question here is what to do with those numbers. You will learn the usual guideline bands, how the four metrics relate, where the bands mislead, and how to turn a set of results into a decision.
The guideline bands
The most widely used thresholds come from the AIAG Measurement Systems Analysis manual. They apply to the gage R&R (GRR) row of the output.
| Metric | Acceptable | Marginal | Unacceptable |
|---|---|---|---|
| %Study variation (GRR sd / total sd) | under 10% | 10% to 30% | over 30% |
| %Contribution (GRR variance / total variance) | under 1% | 1% to 9% | over 9% |
| %Tolerance (6 × GRR sd / tolerance) | under 10% | 10% to 30% | over 30% |
| Number of distinct categories (ndc) | 5 or more | under 5 |
Two of these rows are the same test on different scales. %Study variation is a ratio of standard deviations, and %contribution is a ratio of variances. Squaring one gives the other:
%contribution = (%study variation)² / 100
So 10% study variation is 1% contribution, and 30% is 9%. This is why the %contribution cutoffs look so small. A system with 30% study variation looks poor on that scale, but it "only" takes 9% of the variance. Standard deviations are in the units of the measurement, which is why many practitioners prefer %study variation for judging.
A worked example
Suppose an ANOVA study returns these variance components, in squared millimetres:
- Repeatability: 0.0100
- Reproducibility (operator plus interaction): 0.0025
- Part-to-part: 1.0000
The tolerance is 6.0 mm.
- GRR variance = 0.0100 + 0.0025 = 0.0125, so the GRR standard deviation is 0.1118.
- Total variance = 0.0125 + 1.0000 = 1.0125.
- %Contribution = 100 × 0.0125 / 1.0125 = 1.23%.
- %Study variation = 100 × √(0.0125 / 1.0125) = 11.1%.
- %Tolerance = 100 × 6 × 0.1118 / 6.0 = 11.2%.
- ndc = ⌊1.41 × 1.0 / 0.1118⌋ = ⌊12.6⌋ = 12.
The verdict is mixed. The ndc is far above 5, so the gage can tell parts apart. All three percentage metrics are just above the 10% line, which puts them in the marginal band. Whether that is acceptable depends on the application, which the next sections cover.
How the metrics relate
Each metric answers a different question, so they can disagree.
- %Contribution and %study variation ask what share of the observed spread is measurement noise. Both depend on how much the chosen parts vary.
- %Tolerance asks how much of the specification width the measurement noise uses up. It depends on the tolerance, not on the parts.
- ndc asks how many groups of parts the gage can reliably separate. It is a restatement of the part-to-GRR ratio.
The ndc is tied to %study variation. Let f be the GRR share of total standard deviation. Then ndc = 1.41 × √(1/f² − 1). For f = 0.30 this gives 4.5, which truncates to 4. For f = 0.27 it gives about 5.0. So the ndc ≥ 5 rule is almost the same as "%study variation below about 27%". It adds little information when the parts are representative, and its main use is as a plain-language check.
Because ndc is truncated with a floor, 4.9 reports as 4. Look at the unrounded value before declaring a failure.
The study variation is only as good as your parts
%Contribution and %study variation use total variation estimated from the study's own parts. This creates two failure modes.
- Parts too similar. If you sampled ten parts from a narrow range, part variance is small, and %study variation looks bad even for a precise gage.
- Parts too spread out. If you deliberately included parts at both extremes of the spec, part variance is inflated and the gage looks better than it will in production.
When the study parts do not represent the real process, compare the GRR standard deviation with the process standard deviation from a control chart or capability study. Minitab calls this %process. It answers the question you care about: how much noise does the gage add to the spread I see in production?
%Tolerance avoids this problem. It is also the right metric when the gage is used to sort good from bad parts. If the process is much narrower than the tolerance, a gage with a mediocre %study variation can still be fine for acceptance decisions, because few parts fall near the limits.
The tolerance multiplier also varies. Minitab and most current software use 6 standard deviations, which covers 99.73% of a normal distribution. Older AIAG editions used 5.15, covering 99%. Check which one your software uses before comparing results across tools.
Choosing the metric for the purpose
| Use of the gage | Primary metric | Reason |
|---|---|---|
| Accept or reject parts against a specification | %Tolerance | Error near the limits causes wrong decisions |
| Monitor or control a process, or study its capability | %Study variation against process sd | Noise hides real process shifts and inflates the apparent sd |
| Compare gages or study designs on the same parts | %Contribution or %study variation | Same parts make the ratios comparable |
| Quick check that the gage separates parts | ndc | Easy to explain |
Report all of them in the study record. A system that is acceptable on one and marginal on another is a normal finding, and the purpose of the gage decides which one governs.
A decision flow
Encoding the bands
A small helper makes the bands explicit and keeps the classification consistent across studies. It takes the percentages you already computed.
def classify(metric_pct):
"""Return the AIAG band for a %study variation or %tolerance value."""
if metric_pct < 10:
return "acceptable"
if metric_pct <= 30:
return "marginal"
return "unacceptable"
def verdict(pct_sv, ndc, pct_tol=None):
report = {"study variation": classify(pct_sv),
"ndc": "acceptable" if ndc >= 5 else "unacceptable"}
if pct_tol is not None:
report["tolerance"] = classify(pct_tol)
return report
print(verdict(pct_sv=11.1, ndc=12, pct_tol=11.2))This prints {'study variation': 'marginal', 'ndc': 'acceptable', 'tolerance': 'marginal'}. Keep the thresholds in one place, because some organisations adopt stricter internal limits, such as 20% for critical characteristics.
Handling marginal results
A marginal result is a decision, not a pass or a fail. Weigh these factors:
- Consequence of a wrong call. For a safety-critical dimension, treat 15% tolerance as a problem. For a cosmetic check, 25% may be fine.
- Process capability. If the process is centred and capable with a large margin, measurement error rarely moves a part across a limit.
- Cost of improvement. A new gage may cost far more than the scrap and rework the error causes.
- Which component dominates. Reproducibility problems are often cheap to fix with training and a clearer procedure. Repeatability problems usually need hardware or fixture changes.
Record the decision and the reasoning. Auditors expect a justification for accepting a marginal system.
Gotchas in interpretation
- Truncated negative components. Sampling error can produce a negative variance estimate, which the ANOVA code sets to zero. A zero reproducibility does not mean operators agree perfectly. It means the study could not detect a difference.
- Interaction term. Many packages test the operator-by-part interaction at α = 0.25 and drop it if it is not significant, pooling its sum of squares with repeatability. That changes the GRR slightly. Know which model your software fitted.
- Sampling uncertainty. With 10 parts, 3 operators and 2 replicates, the GRR estimate has few degrees of freedom, and its confidence interval is wide. A result of 9.5% could plausibly be 14% on a repeat. Treat results near a threshold as ambiguous and consider more parts or replicates.
- One study, one condition. The result describes the gage, operators, parts and environment of that day. It does not cover drift over weeks, which is a stability question.
- Thresholds are conventions. The 10% and 30% lines are rules of thumb, not statistical tests. No p-value says a gage is acceptable.
Exercise
A study reports a GRR standard deviation of 0.2, a part standard deviation of 0.9 and a tolerance of 4.0. Compute %study variation, %tolerance (6 sigma) and ndc, then classify each.
Show answer
Total sd = √(0.2² + 0.9²) = √0.85 = 0.922. %Study variation = 100 × 0.2 / 0.922 = 21.7%, which is marginal. %Tolerance = 100 × 6 × 0.2 / 4.0 = 30%, which sits on the boundary of marginal and unacceptable, so treat it as marginal at best and likely a problem for sorting parts. ndc = ⌊1.41 × 0.9 / 0.2⌋ = ⌊6.35⌋ = 6, which meets the 5 or more guideline. The gage separates parts well but uses a large part of the tolerance.
Exercise
A gage shows 12% study variation when ten parts are picked from a narrow band near the nominal value. A colleague wants to reject it. What would you check before agreeing?
Show answer
Check whether the parts represent the real process spread. Narrow part selection shrinks part variance and inflates %study variation. Compare the GRR standard deviation with the process standard deviation from a control chart (%process), and look at %tolerance. If the process sd is much larger than the study part sd, the gage may be fine. If the parts really are representative, 12% is marginal and the decision depends on the application and risk.