Sub topic 13
Resistance from Voltmeter and Ammeter Readings
Prerequisites
Assume you know current, potential difference, Ohm’s law, series and parallel circuits, and percentage error. A resistance is found from
\[ R=\frac{V}{I} \]
where \(V\) is the potential difference across the resistor and \(I\) is the current through it.
Circuit arrangement
An ammeter measures current, so it is connected in series with the resistor. A voltmeter measures potential difference, so it is connected in parallel across the resistor.
An ideal ammeter has zero resistance. An ideal voltmeter has infinite resistance. Real instruments are not ideal:
- The ammeter has a small internal resistance \(r_A\).
- The voltmeter has a large but finite resistance \(R_V\).
These resistances can affect the measured value.
Finding resistance from readings
Take several pairs of readings by changing the rheostat setting. For each pair, calculate \(R=V/I\), or plot \(V\) against \(I\).
For an ohmic resistor, the graph is a straight line through the origin:
\[ V=IR \]
Therefore, the slope of a \(V\)-against-\(I\) graph is the resistance:
\[ R=\frac{\Delta V}{\Delta I} \]
Using a best-fit line is better than using one pair of readings because it reduces the effect of random fluctuations.
I = [0.1, 0.2, 0.3, 0.4] # current in A
V = [0.52, 0.98, 1.53, 2.01] # potential difference in V
n = len(I)
mean_I = sum(I) / n
mean_V = sum(V) / n
slope = sum((x - mean_I) * (y - mean_V)
for x, y in zip(I, V)) / sum((x - mean_I) ** 2 for x in I)
print(f"R = {slope:.2f} ohm")Instrument loading and error
There are two common ways to connect the meters.
| Arrangement | Main effect | Suitable when |
|---|---|---|
| Voltmeter across resistor and ammeter outside the branch | Ammeter measures resistor current plus voltmeter current | The resistor has high resistance |
| Voltmeter across resistor and ammeter inside the branch | Ammeter measures only resistor current, but voltmeter also measures the ammeter’s voltage drop | The resistor has low resistance |
In the first arrangement, the measured resistance is slightly smaller than the actual resistance because the ammeter reads too much current. In the second arrangement, it is slightly larger because the voltmeter reads the resistor voltage plus the voltage drop across the ammeter.
Choose the arrangement that makes the meter-induced error small. A high-resistance voltmeter and a low-resistance ammeter reduce this error.
Error in a resistance reading
If \(R=V/I\), the maximum fractional error is approximately
\[ \frac{\Delta R}{R}=\frac{\Delta V}{V}+\frac{\Delta I}{I} \]
Use readings that are large enough to make the percentage errors small, while staying within the safe range of both meters.
Exercise
A resistor has readings V = 3.0 ± 0.1 V and I = 0.50 ± 0.02 A. Find its resistance and approximate maximum percentage error.
Show answer
R = 3.0/0.50 = 6.0 ohm. Percentage error = (0.1/3.0 + 0.02/0.50) × 100% ≈ 7.3%.