Sub topic 1
Measurement, Units and Precision in Experimental Physics
Prerequisites
Assume basic units and dimensions of length, simple decimal arithmetic, and the idea that a measurement compares an unknown quantity with a standard unit.
Measurement and units
A measurement has a numerical value and a unit. For example, 2.36 cm means a length 2.36 times the centimetre standard. The SI unit of length is the metre, but millimetres and centimetres are convenient for laboratory instruments.
Instruments differ in precision because their smallest readable interval differs. A ruler may read to 1 mm, while a vernier calliper may read to 0.01 cm. A smaller interval allows finer readings, but it does not remove all errors.
Least count
The least count is the smallest change in a quantity that an instrument can measure reliably.
| Instrument | Typical least count | Common use |
|---|---|---|
| Metre scale | 1 mm | Ordinary length |
| Vernier calliper | 0.01 cm | Diameter, depth, length |
| Screw gauge | 0.01 mm | Thin wire or sheet |
The least count sets the appropriate precision of the final answer. Do not report more digits than the instrument supports.
Errors, correction and significant figures
A systematic error has a consistent cause and shifts readings in one direction. A random error changes unpredictably between readings. Repeating measurements and taking their mean reduces random error, but not an uncorrected systematic error.
Zero correction removes the reading shown when the true quantity is zero:
Correction = − zero error
Use significant figures that match the instrument’s precision. The digits that carry meaningful measurement information are called significant figures.
Vernier calliper: principle and least count
A vernier calliper has a main scale and a sliding vernier scale. Its principle is that a chosen number of vernier divisions is slightly different in length from the same number of main-scale divisions. This small difference permits finer measurement.
For a direct vernier:
Least count = 1 MSD − 1 VSD
If 10 VSD = 9 MSD and 1 MSD = 0.1 cm, then:
LC = 0.1/10 = 0.01 cm
Reading and correcting a vernier calliper
- Record the main scale reading just to the left of the vernier zero.
- Find the vernier division that exactly coincides with a main-scale mark.
- Calculate the vernier contribution: coinciding division × least count.
- Add it to the main scale reading.
- Subtract the signed zero error.
For MSR = 2.3 cm, the sixth division coinciding, LC = 0.01 cm, and zero error = +0.02 cm:
Observed reading = 2.3 + 6(0.01) = 2.36 cm
Corrected reading = 2.36 − 0.02 = 2.34 cm
A positive zero error occurs when the vernier zero lies to the right of the main zero with closed jaws. A negative zero error occurs when it lies to the left. Apply the opposite correction.
Exercise
A calliper has 1 MSD = 0.1 cm and 10 VSD = 9 MSD. The main reading is 1.8 cm, the fifth vernier division coincides, and the zero error is −0.01 cm. Find the corrected reading.
Show answer
LC = 0.01 cm, observed reading = 1.8 + 5(0.01) = 1.85 cm. Corrected reading = 1.85 − (−0.01) = 1.86 cm.