Moving coil galvanometer and its conversion to ammeter and voltmeter
This chapter assumes you know the torque on a current loop, τ = NIAB sin θ, and the series and parallel rules for resistors. It uses the magnetic dipole moment m = NIA from the previous chapter.
How the galvanometer works
A moving coil galvanometer detects small currents. A rectangular coil of N turns and area A hangs between the curved pole pieces of a permanent magnet. The pole pieces and a soft iron core make the field radial. In a radial field, the plane of the coil stays parallel to B at every angle of rotation, so the magnetic torque is always
τ_magnetic = NIAB
and does not depend on the deflection angle. A spring provides a restoring torque kφ, where k is the torsional constant and φ is the deflection. The coil stops when the two torques balance:
NIAB = kφ, so φ = (NAB/k) I
The deflection is proportional to the current, so the scale is linear. The ratio current sensitivity is φ/I = NAB/k. The ratio voltage sensitivity is φ/V = NAB/(kR), where R is the coil resistance.
Increasing sensitivity
To increase current sensitivity, increase N, A or B, or use a spring with a smaller k. Increasing N also increases the coil resistance R. Voltage sensitivity depends on N/R, so it does not always improve when N grows. If N doubles, R roughly doubles, and the voltage sensitivity stays about the same.
Converting to an ammeter
An ammeter measures large currents and must have a very small resistance, because it sits in series with the circuit. Connect a small shunt resistance S in parallel with the galvanometer. The galvanometer has resistance G and full-scale current I_g. To measure a maximum current I:
Ig G = (I − Ig) S, so S = Ig G / (I − Ig)
The combined resistance is GS/(G + S), which is smaller than S.
Converting to a voltmeter
A voltmeter sits in parallel with the part of the circuit it measures, so it must have a very large resistance. Connect a large resistance R in series with the galvanometer. To measure a maximum voltage V:
V = Ig (G + R), so R = V/Ig − G
The total resistance is G + R, which is larger than G.
| Instrument | Added resistor | Connection | Total resistance |
|---|---|---|---|
| Ammeter | Shunt S = I_g G/(I − I_g) | Parallel | Less than S |
| Voltmeter | R = V/I_g − G | Series | More than G |
Worked example
A galvanometer has G = 50 Ω and full-scale current I_g = 2 mA. Find the shunt for a 0–1 A ammeter and the series resistance for a 0–10 V voltmeter.
G = 50.0 # ohm
Ig = 2e-3 # ampere
# Ammeter, range 0 to 1 A
I = 1.0
S = Ig * G / (I - Ig)
r_ammeter = G * S / (G + S)
print(S, r_ammeter) # about 0.1002 ohm, 0.1000 ohm
# Voltmeter, range 0 to 10 V
V = 10.0
R = V / Ig - G
r_voltmeter = G + R
print(R, r_voltmeter) # 4950 ohm, 5000 ohmThe ammeter's resistance is about 0.1 Ω, so it barely disturbs the circuit. The voltmeter's resistance is 5000 Ω, so it draws only 2 mA at full scale.
Common errors
- Using I instead of I − I_g in the shunt formula. The shunt carries only the current the galvanometer does not.
- Forgetting to subtract G in the voltmeter formula.
- Saying that a larger N always raises voltage sensitivity. R also grows with N.
- Forgetting that the radial field keeps the torque equal to NIAB, with no sin θ factor.
Exercise
A galvanometer has G = 100 Ω and shows full-scale deflection for 5 mA. What shunt converts it to an ammeter reading up to 2.005 A, and what series resistance converts it to a voltmeter reading up to 20 V?
Show answer
Shunt: S = Ig G/(I − Ig) = (0.005 × 100)/(2.005 − 0.005) = 0.5/2 = 0.25 Ω, in parallel. Series resistance: R = V/I_g − G = 20/0.005 − 100 = 4000 − 100 = 3900 Ω, in series.
Exercise
The number of turns in a galvanometer coil is doubled, and the other quantities stay the same, except that the coil resistance also doubles. How do the current sensitivity and the voltage sensitivity change?
Show answer
Current sensitivity NAB/k doubles. Voltage sensitivity NAB/(kR) stays the same, because N and R both double.